Description
THC & Methane Analyzer is designed to minimize the loss of hydrocarbons prior to analysis by maintaining the required temperature through use of a single heated filter, single heated sample line and stainless steel fittings, all of which have low hydrocarbon adsorption characteristics. The system has been designed to endure the harsh environment of the in-use test application while maintaining laboratory grade measurement standards.
Feature
- Automatic Flame Ignition: The FID flame will light automatically after approximately 30 minutes of warm-up time.
- Dual Flame Ionization Detector Assembly: The single assembly dual detector system, coupled with an NMHC cutter assembly upstream the CH4 FID, provides both THC and CH4 measurements in real-time using only 1 sample line, 1 FID fuel line and 1 combustion air input.
- Measuring Range: The Dual FID has 4 user-selectable measurement ranges for each channel, ranging from 0-100 ppmC1 and 0-30000 ppmC1.
- Sample Handling: A water trap and Nafion dryer are integrated into the SEMTECH SCS module. A heated sample line (length varies with tested vehicle) delivers the sample to the Dual FID chamber at 191°C. Particulate matter is removed using a built-in, disposable, user-replaceable filtration system.
- Combustion Air: The Dual FID module includes a built-in catalytic converter for removal of all hydrocarbons from ambient air, thus providing hydrocarbon-free combustion air to the system.
- NMHC Cutter: The system includes a built-in NMHC catalytic oxidation assembly for removal of the NMHC fraction from sample prior to analyzing the sample using the Dual FID’s methane channel.
- Compliance: The SEMTECH Dual FID is compliant for gases regulated under the rules of UN-ECE. Reg. 49 and E.U. Commission Regulation No. 582/2011 and of US EPA 40CFR, Part 1065.
- Heated Lines: Sensors, Inc. has developed an innovative, low powered heated line that controls and maintains the temperature of the exhaust sample at 191 ±5°C.
- 12V DC Power: The system operates on 12V DC power.
Spesification
|
Power requirements |
12 V |
|
Warm up Time |
30 minutes minimum; 60 minutes to meet performance specifications |
|
Storage Temperature |
Dry -10°C to 60°C ambient |
|
Operating Environment |
-10°C to 45°C ambient |
|
Power Usage |
|
|
Dimensions (WxDxH) |
|
|
Weight |
|
|
Data transmission |
Ethernet |
|
Electromagnetic interference and susceptibility |
CE Standards: IEC 61326: 2002-2 |
|
Gas |
THC |
|
Supported ranges |
30000 ppmC1 10000 ppmC1 1000 ppmC1 100 ppmC1 |
|
Linearity (all ranges) |
|xmin x (a1 - 1) + a0 | ≤ 0.5% of selected range Slope a1 between 0.99 and 1.01 Standard Error of Estimates (SEE) ≤ 1% of selected range Coefficient of Determination r2 ≥ 0.998 |
|
Accuracy |
≤ ±2% of reading or ≤ ±0.3% of full scale of selected range, whichever is greater |
|
Repeatability |
< ±1% of point (span) or < ±1% of selected range, whichever is greater |
|
Precision |
≤ 1% of selected range |
|
Noise |
≤ 2% of selected range |
|
Zero drift |
≤ ± 1% of full scale of selected range over 1 hour |
|
Span drift |
≤ ± 2% of full scale of selected range over 8 hours |
|
Rise Time T10-90 |
≤ 2.5 seconds |
|
System Response Time T10-90 |
≤ 10 seconds with rise time ≤ 2.5 seconds |
|
Data Rate |
5 Hz |
|
Gas Requirements & Consumption |
|
|
Fuel |
H2 40% Bal. He ( |
|
Fuel consumption |
150 ccm @ 300 kPa |
|
Span gas |
C3H8 |
|
Span gas consumption |
~1.5 LPM @ 150 - 200 kPa |
|
Zero gas |
Purified N2 or Purified Synthectic Air |
|
Zero gas consumption |
~1.5 LPM @ 150 - 200 kPa |
|
Total Sample Flow |
|