Description
GCOM-3000 is a VOCs monitoring product that uses international mainstream chromatographic methods. It can monitor Organic pollutant gases such as methane, the total hydrocarbon, nonmethane total hydrocarbon,low-carbon aldehyde ketone, benzene series (benzene, toluene, xylene, ethylbenzene, isopropylbenzene) and part of halogenated hydrocarbons by using FID detector and high-performance chromatographic separation column. The pneumatic heating box is designed inside the analyzer which can solve the problem of the measured results is low under the certain high-temperature conditions.
Feature
- After each sample injection, the analyzer automatically blows the chromatographic column to speed up the component sieving and analysis;
- The gas line is inertly processed to avoid the adsorption and residue of samples on the pipe wall;
- The software in addition to the built-in default analysis method, also allows the user to free configure analysis process and methods;
- The configuration supports multiple detectors which can be used for complex sample analysis at the same time and can improve the ability of analysis;
- Support TCP/IP network remote control.
Spesification
- Method: GC-FID
- Range: 0.05ppm~1000ppm(CH4)
- Distinguishability: 0.001ppm
- Qualitative Repeatability: RSD7≤1.0%
- Quantitative Reproducibility: RSD7≤2.0%
- Range Drift: ≤3.0%(4h)
- Oven temperature stability: ≤0.5oC
- Meas.Period: 2mins~30mins (the time will change according to
- different component separation columns)
- Dimension: 483mm×266mm×560mm
- Display: touch LCD screen
- Comm. Output: 10/100M ethernet port , RS232
- Comm. Protocol: Modbus-RTU/TCP
- Power Supply: 220VAC, 50Hz~60Hz, 900W