GCOM-3000 Volatile Organic Compounds Online Analyzer

GCOM-3000 Volatile Organic Compounds Online Analyzer



Order

Description

GCOM-3000 is a VOCs monitoring product that uses international mainstream chromatographic methods. It can monitor Organic pollutant gases such as methane, the total hydrocarbon, nonmethane total hydrocarbon,low-carbon aldehyde ketone, benzene series (benzene, toluene, xylene, ethylbenzene, isopropylbenzene) and part of halogenated hydrocarbons by using FID detector and high-performance chromatographic separation column. The pneumatic heating box is designed inside the analyzer which can solve the problem of the measured results is low under the certain high-temperature conditions.
 

Feature

  • After each sample injection, the analyzer automatically blows the chromatographic column to speed up the component sieving and analysis;
  • The gas line is inertly processed to avoid the adsorption and residue of samples on the pipe wall;
  • The software in addition to the built-in default analysis method, also allows the user to free configure analysis process and methods;
  • The configuration supports multiple detectors which can be used for complex sample analysis at the same time and can improve the ability of analysis;
  • Support TCP/IP network remote control.

Spesification

  • Method: GC-FID
  • Range: 0.05ppm~1000ppm(CH4)
  • Distinguishability: 0.001ppm
  • Qualitative Repeatability: RSD7≤1.0%
  • Quantitative Reproducibility: RSD7≤2.0%
  • Range Drift: ≤3.0%(4h)
  • Oven temperature stability: ≤0.5oC
  • Meas.Period: 2mins~30mins (the time will change according to
  • different component separation columns)
  • Dimension: 483mm×266mm×560mm
  • Display: touch LCD screen
  • Comm. Output: 10/100M ethernet port , RS232
  • Comm. Protocol: Modbus-RTU/TCP
  • Power Supply: 220VAC, 50Hz~60Hz, 900W